A biosimilar does not have to repeat the full development of the reference product. It has to show that it is highly similar in quality, with no clinically meaningful differences in safety and efficacy. The evidence for that is built step by step, and the analytical comparison carries most of the weight.
The stepwise approach
- Analytical similarity: structure, physicochemical properties and biological activity compared with several batches of the reference product.
- Non-clinical data: mainly in-vitro functional assays; animal studies are now rarely expected.
- Clinical data: pharmacokinetic and, where needed, pharmacodynamic comparison. Agencies increasingly accept that a comparative efficacy trial may not be needed when the analytical and PK data are strong.
What the analytical package must show
- Primary structure and higher-order structure.
- Glycosylation, charge variants, size variants and other product-related substances and impurities.
- Biological activity through assays that reflect the mechanism of action, including Fc-related functions for antibodies.
- Enough reference product batches, over time, to set meaningful similarity ranges.
- A justification for every difference found, showing that it has no clinical impact.